P1 Patient Monitor User’s Manual
7-14
Continuous: The nurse call signal lasts until the alarm ends. That is
to say the duration of a nurse call signal is equal to that of the alarm
condition
.
4. Select Trigger Typeto set the work mode of the nurse call relay.
5. Select Alarm Priorityto set the priority of alarms sent to the nurse call
system.
6. Select Alarm Typeto set the type of alarms sent to the nurse call system.
WARNING
Do not rely exclusively on the nurse call system for alarm notification.
Remember that the most reliable alarm notification combines audible and
visual alarm indications with the patient’s clinical condition.
7.12. Intubation Status
The monitor provides the intubation status function during RESP and CO
2
monitoring. In this state, the physiological alarms related to RESP and CO
2
are shielded,
and the alarm off icon is displayed in the parameter area. During the intubation process of
general anesthesia surgery, the intubation status can be selected to shield unnecessary
alarms.
7.12.1. Entering Intubation Status
To enter the intubation status, choose either of the following ways:
SelectIntubation Statusquick key.
From the bottom of the RESP or CO
2
menu to select Intubation
Status button.
Select Main Menu quick keyfrom Alarm column to select
Intubation Status.