P1 Patient Monitor User’s Manual
2
A.2 Environmental Specifications
Components Item Operating condition Storage condition
Mainframe and
other modules
Temperature (℃) 0~40 -20~+60
Relative humidity
(non-condensing) (%)
15~95 10~95
Barometric (kPa)
54.0~107.4 16.0~107.4
Sidestream CO
2
(C1)
Temperature (℃) 5~40 -40~+70
Relative humidity
(non-condensing) (%)
10~90
<90
Barometric (kPa)
55.0~107.4 55.0~107.4
CAUTION:
The equipment must be used under the specified environmental specifications,
otherwise it will not meet the technical specifications claimed in this manual
and may lead to unexpected consequences such as equipment damage. If the
performance of the equipment changes due to aging or environmental
conditions, please contact the maintenance personnel.
A.3 Physical Specifications
Parts Model Weight
Size (H×L×D)
Remark
Main unit P1 <0.9kg 146 ×99 ×76 /
Transfer Handle TH1 <0.4kg 146 ×165 × 81 /
Docking station DS1 <0.9kg 182× 165 × 123 /
SpO
2
module S2 <0.3kg 136.6×102×40 Nellcor SpO
2
SpO
2
module S5 <0.3kg 136.6×102×40 Masimo SpO
2
Mainstream CO
2
module
C8 <0.3kg 136.6×102×40
BLT CO
2
Sidestream CO
2
module
C1 <0.4kg 136.6×102×40
BLT CO
2
NIBP module N2 <0.3kg 136.6×102×40 SunTech NIBP