P1 Patient Monitor User’s Manual
7-5
7.7. Setting Alarm
You can set the alarm properties centrally. Select【Alarm Setup】quick keyor select
the corresponding button from the【Alarm】column,
in the main menu to set alarm.
7.7.1. Setting Parameter Alarm
The steps to centrally set the properties of the parameter alarm are as follows:
1. Enter【Limit】interface in any of the following ways:
Select 【Alarm Setup】 quick key.
Select 【Main Menu】 quick keyfrom【Alarm】column to select【Limit】.
2. Select parameter submenu to set the alarm according to the required. You can
also set the alarm for individual parameters from the parameters menu.
7.7.2. Changing Alarm Setup Protection Mode
You can change the password protection mode of the alarm settings and arrhythmia
settings as follows:
1. Select 【 Main Menu 】 quick keyfrom 【 System 】 column to select
【Maintenance】input maintenance passwordEnter.
2. Select【Authorization】submenu.
3. Change the password protection mode of the alarm settings.
【No Password】: Change alarm setups to not be password protected.
【Password】: Change alarm switch, alarm limit and alarm level to be
protected by password.
If you use password to access alarm and arrhythmia alarm related settings, you can
set the valid time of the password, beyond which you need to re-enter the password. For
details, please refer to 5.9 Setting Password Valid Time.
7.7.3. Setting Alarm Sound Properties
7.7.3.1. Setting Alarm Volume
1. Enter【Setup】interface in either of the following ways: