P1 Patient Monitor User’s Manual
7-10
When the CMS/eGW Disconnected switch is turned on, the technical
alarm will be generated when the monitor and the CMS/eGateway are
disconnected after successful connection.
7.8. Alarm Pause
When the alarm is paused, it has the following characteristics:
Shield all physiological alarms within the set time.
The technical alarm sound is paused, but the alarm light and alarm
information are still displayed.
Display the remaining time of alarm paused in the physiological alarm
information area.
Display the alarm paused icon in the information area.
After reaching the alarm pause time, the monitor will automatically exit the alarm
pause state. You can also click Alarm Pausequick key to manually cancel the alarm
pause.
7.8.1. Setting Alarm Pause Time
Alarm pause time can set to :1min, 2min, 3minandPermanent, the
default is 2 minutes. The steps to set the alarm pause time are as follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenu Pause /Resetsubmenu.
3. SetAlarm Pause Duration.
7.8.2. Turn off all the Alarm
If Alarm Pause Durationis set to Permanent(Refer to section 7.8.1 Setting
Alarm Pause Time), you can pressAlarm Pausequick key to turn off all alarms.
When
the alarm is turned off, it has the following characteristics:
No physiological alarm lamps flash and no physiological alarms are sounded.
The technical alarm sound is turned off, but the alarm light and alarm
information are still displayed
Display "Alarm Off" in the physiological alarm information area and the