P1 Patient Monitor User’s Manual
39
C.2 RESP Default Settings
Item Default Setting
RR Alarm switch ON
High limit Adult / Pediatric: 30
Neonatal: 100
Low limit Adult / Pediatric: 8
Neonatal: 30
Alarm priority MED
Alarm print OFF
Apnea Alarm switch ON
Alarm priority HIGH, unadjustable
Alarm print OFF
Apnea Delay 20 s
RR Source Auto
RESP Lead RA_LL
Gain ×1
Wave Speed 6.25 mm/s
Auto Threshold Detection ON
Respiratory anti-drift ON
C.3 SpO
2
Default Settings
Item Default Setting
SpO
2
Alarm switch ON
High limit
Adult / Pediatric:100
Neonatal:95
Low limit
90
Alarm priority MED
Alarm print OFF
Desat Alarm switch ON