P1 Patient Monitor User’s Manual
11-3
11.5.3. Displaying the Temperature Difference
To display the temperature difference between two measurement sites monitored by
the same temperature module, switch on corresponding T. To do so, follow this
procedure:
1. Select the TEMP parameter area to enter the TEMPmenu.
2. SelectSetupsubmenu.
3. Switch on 󰅿T.
11.5.4. Setting TEMP Unit
You can change the unit of TEMP by following the steps below:
1. Select the TEMP parameter area to enter theTEMPmenu.
2. SelectSetupsubmenu.
3. Set TEMPUnit.
11.6. TEMP Troubleshooting
This section lists the problems that might occur. If you encounter the problems when
using the equipment or accessories, check the table below before requesting for services.
If the problem persists, contact your service personnel.
NOTE
:
For the physiological and technical alarm messages, see D Alarm
Message.
Problem Corrective Actions
Do not display TEMP parameter
area on the main screen.
1. Check if the display of the TEMP parameter is set in
Screen Setupmenu.
2. Check that if the TEMP parameter switch is enabled.
For more information, see 3.5.1 Setting Parameters.
3. Check that if the connections of the temperature
probe and the extension cable are tight.
Measurement fails / "--" is 1. If you are using a disposable probe, check whether