P1 Patient Monitor User’s Manual
7-11
background color is red.
Display alarm off icon in status area.
To exit the alarm off state, click【Alarm Pause】quick key
again.
WARNING:
Pausing or off the alarm may cause the patient to be in danger, please handle it
carefully.
7.9. Alarm Reset
Click on【Alarm Reset】quick key to reset the alarm system, and the alarm reset icon
will appear in the system status information area.
NOTE
:
In the alarm reset state, if a new alarm is generated, the alarm reset icon
disappears and the alarm system is reactivated.
7.9.1. Physiological Alarm Reset
After the physiological alarm is reset, the sound of the currently existing
physiological alarm is shielded, and the other alarm states remain unchanged.
7.9.2. Technical Alarm Reset
When the technical alarm is reset, it has the following characteristics:
The technical alarm that can be completely cleared is cleared. The monitor
will not have any alarm indication for the cleared technical alarm.
Technical alarm that can clear sound and light is displayed as prompt
message.
The sound of the technical alarm that cannot be cleared is shielded. For the
indication of the technical alarm after the alarm is reset, please refer to D.2
Technical Alarm.