P1 Patient Monitor User’s Manual
7-16
button in the Alarm submenu will be activated. Click on Reset Remote Alarm
button,
other bed alarm will be reset.
CAUTION
Only when the “Alarm Reset By Other Bed” function of the remote monitor is
enabled, can you reset other bed alarm on this monitor
7.13.2. Authorizing the Alarm Reset to Other Devices
Alarms on your monitor can be reset by remote devices if you enable this function.
To do so, follow this procedure:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuRemote Viewsubmenu.
3. OpenAlarm Reset By Other Bedswitch.
7.13.3. Switching Off the Remote Device Disconnection Alarm
The monitor can provide an alarm if remote devices are disconnected. By default, the
function is enabled. To disable the alarm, follow this procedure:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuRemote Viewsubmenu.
3. Switch offRemote Disconnected Alarm.
7.14. Detecting Alarm
The monitor automatically performs a self-test at startup. Check that the alarm lamp
illuminates, one after the other, in red, yellow, and cyan, and that an alarm tone is heard.
This indicates that the visible and audible alarm indicators functions correctly.
To further test individual measurement alarms, perform measurements on yourself or
using a simulator. Adjust alarm limits and check that appropriate alarm behavior is
observed