adjusts the RESP waveform threshold. For more information, see
9.5.7.Manually Adjust the RESP Waveform Detection Threshold.
9.5.7. Manually Adjust the RESP Waveform Detection Threshold
Use the manual detection mode in the following situations:
The respiration rate and the heart rate are close.
Patients have intermittent mandatory ventilation.
Respiration is weak. Try repositioning the electrodes to improve the signal.
To set the Resp waveform threshold to the desired level, follow this procedure:
1. Select the RESP parameter area or waveform area to enter the 【RESP】menu.
2. Select【Threshold】submenu.
3. Select
the up and down arrows below【Threshold】to define the Resp
waveform threshold.
Once set, the detection level will not adapt automatically
to different respiration depths. It is important to remember that if the depth of
breathing changes, you may need to change the detection level.
9.5.8. Setting the Respiration Filter
Turn on the respiration filter function can filter out the interference in the respiration
waveform. The steps to set the respiration filter switch are as follows:
1. Select the RESP parameter area or waveform area to enter the 【RESP】menu.
2. Select【Setup】submenu.
3. Turn on or turn off【RESP Filter】.
9.6. RESP Troubleshooting
For more information, see D Alarm Message.