P1 Patient Monitor User’s Manual
45
Item Default Setting
Scale Type Manual
Scale Upper scale ART /Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure:
160mmHg
CVP /ICP/RAP/LAP/UVP venous pressure : 20mmHg;
PA/PAWP30mmHg
Lower scale 0mmHg
High-precision cursor
switch
OFF
High-precision cursor ART/ Ao/ UAP/BAP/FAP/LV/P1/P2 arterial pressure:
80mmHg
CVP /ICP/RAP/LAP/UVP venous pressure: 10mmHg
PA/PAWP15mmHg
PPV measurement OFF
PPV source Auto
Overlapping
IBP
Waveforms
Left Scale 0~160mmHg
Right Scale P1/P2: 0~160mmHg
CVP/RAP/LAP/ICP/UVP: 0 ~20 mmHg
Wave speed 25 mm/s
Grid OFF
C.7 CO
2
Default Settings
Item Default Setting
EtCO
2
Alarm switch ON
High limit Adult/Pediatric: 50 mmHg
Neonatal: 45 mmHg
Low limit Adult/Pediatric: 25mmHg
Neonatal: 30mmHg
Alarm
priority
MED