P1 Patient Monitor User’s Manual
7-13
CAUTION:
Changing of alarm priority may affect the latching status of corresponding
alarm. Please determine if you need to reset the latching status for the specific
alarm when you have changed its alarm priority.
When the alarm system is reset, the latched physiological alarms are cleared.
Do not set
all alarm status to latching alarm signals when used in the intensive
care unit.
7.11. Nurse Call
The nurse call function means that when the alarm set by the user occurs, the monitor
can output a signal to the nurse call system, call the nurse. The monitor provides a nurse
call connector, and the monitor is connected to the nurse call system of the hospital
through the randomly provided nurse call cable. After the system is connected, the
connector can implement the nurse call function.
The nurse call function must be valid only if the following conditions are met:
The nurse call function is turned on.
A user-defined alarm occurs.
The monitor is not alarm paused or off.
7.11.1. Changing Nurse Call Settings
To set the type and priority of alarms that are sent to the nurse call system, follow this
procedure:
1. Select 【 Main Menu 】 quick keyfrom 【 System 】 column to select
【Maintenance】input maintenance passwordEnter.
2. Select【Alarm】submenu【Nurse Call】submenu.
3. Select【Signal Type】to set the type of nurse call signal.
【Pulse】: The nurse call signal is a pulse signal and each pulse lasts one
second. When multiple alarms simultaneously occur, only one pulse
signal is outputted. If an alarm occurs but the previous one is not cleared,
a new pulse signal will also be outputted.