P1 Patient Monitor User’s Manual
19-2
CAUTION:
If you accidentally pour liquid on the monitor, plug-in modules or accessories,
please contact the maintenance personnel or our company immediately. Please
do not use the equipment until it has been detected and confirmed that it can
continue to be used.
To clean or disinfect reusable accessories, please refer to the instructions
provided with the accessories.
19.2. Cleaning the monitor and other mounting accessories
The monitors, plug-in modules, docking station and transfer handles should be
cleaned regularly. If there is heavy pollution or lots of dust and sand in your place, the
frequency of cleaning should be increased. Before cleaning, please consult the hospital’s
regulations for cleaning.
Use a soft cloth that cannot bear balls to wet and clean the surface with an
appropriate amount of water or alcohol-based detergent (such as 70% ethanol), and avoid
interfaces and metal parts of equipment. Do not use strong solvents such as acetone or
tichlorothylene. Be careful when cleaning the monitor’s screen, which is more sensitive
than the case. After cleaning, wipe the cleaner off the surface of the mainframe and other
mounting accessories with a dry cloth, and place it in a ventilated and cool environment to
dry.
CAUTION
Interfaces and metal parts may be corroded after contacting with detergent.
19.3. Disinfecting the monitor and other mounting accessories
You can disinfect the monitor, plug-in modules, docking station and transfer handles
according to the hospital’s disinfection procedures. The equipemnts should be cleaned
before disinfection. The following table lists the recommended disinfectants: