P1 Patient Monitor User’s Manual
42
Item Default Setting
Neonatal: 25 mmHg
Alarm priority MED
Alarm print OFF
NIBP-sdp Alarm switch ON
High limit 60mmHg
Low limit 20mmHg
Alarm priority MED
Alarm print OFF
Initial Pressure Adult: 160 mmHg
Pediatric: 130 mmHg
Neonatal: 100 mmHg
Interval manual
Start Mode Clock
NIBP End Tone OFF
Venipuncture pressure Adult: 80mmHg
Pediatric: 60mmHg
Neonatal: 40mmHg
Unit mmHg
C.6 IBP Default Settings
Item Default Setting
IBP-S Alarm switch ON
High limit ART/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure
Adult: 160 mmHg
Pediatric: 120 mmHg
Neonatal: 90 mmHg
PA/PAWP:
Adult: 35 mmHg
Pediatric/Neonatal: 60mmHg
Low limit ART/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure