P1 Patient Monitor User’s Manual
7-12
7.10. Latching Alarms
The physiological alarms are classified into “Latching” and “Non-latching”.
Non-latching alarms:
After the condition that triggered the alarm of a
parameter disappears, the system will not make any prompt for this alarm of
this parameter.
Latching alarms: Even if the condition that caused the physiological alarm
disappears, the alarm signal will still be “Latched”, and the time of the last
triggering of the alarm will be displayed behind the alarm information in the
information area.
You can choose to individually lock the visual signal or simultaneously lock
the visual and audible signals.
For visual latching, after the alarm condition disappears, the visual signal of
the alarm, including the alarm light, the alarm information and the
background color remain unchanged, and the alarm information text is
followed by the time of last triggering the alarm.
For audible latching, the system still emits an alarm tone after the alarm
condition disappears.
The steps to latch the physiological alarm are as follows:
1. Select 【 Main Menu 】 quick keyfrom 【 System 】 column to select
【Maintenance】input maintenance passwordEnter.
2. Select【Alarm】submenu【Latching】submenu.
3. Select how you want to latch the alarms. Alarm latching rules are as follows:
You can separately select
visual latching.
Latching audible alarm signal simultaneously latches visual signal
corresponding to the alarm level.
When a low priority alarm is latched, the high priority alarm is also
automatically locked. For example, if you select the low priority alarm,
the medium priority alarm and the high priority alarm will also be
latched simultaneously.