P1 Patient Monitor User’s Manual
41
Item Default Setting
Alarm priority MED
Alarm print OFF
Unit ºC
C.5 NIBP Default Settings
Item Default Setting
NIBP-S Alarm switch ON
High limit Adult: 160 mmHg
Pediatric: 120 mmHg
Neonatal: 90 mmHg
Low limit Adult: 90 mmHg
Pediatric: 70 mmHg
Neonatal: 40 mmHg
Alarm priority MED
Alarm print OFF
NIBP-D Alarm switch ON
High limit Adult: 100 mmHg
Pediatric: 70 mmHg
Neonatal: 60 mmHg
Low limit Adult: 60 mmHg
Pediatric: 40 mmHg
Neonatal: 20 mmHg
Alarm priority MED
Alarm print OFF
NIBP-M Alarm switch ON
High limit Adult: 115 mmHg
Pediatric: 90 mmHg
Neonatal: 70 mmHg
Low limit Adult: 70 mmHg
Pediatric: 50 mmHg