P1 Patient Monitor User’s Manual
12-10
this procedure:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectModulesubmenuOthersubmenu.
3. SetNIBP Invalid Time.
12.9.9. Displaying the NIBP List
To display multiple sets of the latest NIBP measurements, follow this procedure:
1. EnterScreen Layoutsubmenu
by either of the following ways:
SelectScreen Setupquick keySelectScreen Layoutsubmenu.
SelectMain Menuquick keyfrom Display column to select
Screen Layout.
2. In the desired parameter area, selectNIBPNIBP
List.
12.10. Assisting Venous Puncture
You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood
vessel and therefore help venous puncture. To assist venous puncture, follow this
procedure:
1. Select the NIBP parameter area.
2. SelectSetupsubmenu
3. SetVenipuncture Pressure.
4. SelectVenipunctureat the bottom of the menu.
5. Puncture vein and draw blood sample.
6. SelectNIBP Start/Stopquick key orVenipuncturebutton to manually
deflate the cuff. When performing a venipuncture, observe the inflation pressure and the
remaining time of the venipuncture in the NIBP parameter area.
12.11. NIBP Maintenance
12.11.1. NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. The NIBP