P1 Patient Monitor User’s Manual
12-9
Interval: After the first measurement, the monitor automatically
repeats measurements at set interval. For example, if Interval is set
to30min, and you start NIBP auto measurement at 10:03, the next
measurement will be taken at 10:33, and then at 11:03, 11:33, and so on.
12.9.5. Enabling the NIBP End Tone
The monitor can issue a reminder tone at the completion of NIBP measurement. The
NIBP End Tone is off by default. To switch on the NIBP end tone, follow this procedure:
1. Select the NIBP parameter area a to enter theNIBPmenu
2. SelectSetupsubmenu.
3. Switch onNIBP
End Tone.
12.9.6. Setting NIBP Sequence Measurement
NIBP sequence measurements can consist of up to 5 measurement periods: A, B, C,
D, and E. You can set the measurement duration for each period and the interval between
NIBP measurements in each period separately. The steps to set up the NIBP measurement
sequence are as follows:
1. Select the NIBP parameter area a to enter theNIBPmenu.
2. SelectSequencesubmenu.
3. Set each sequence measurement toDurationorInterval Timeseparately.
12.9.7. Setting NIBP Unit
You can change the NIBP units by following these steps:
1. Select the NIBP parameter area a to enter theNIBPmenu.
2. SelectSetupsubmenu.
3. Set NIBPUnit.
12.9.8. Setting NIBP Invalid Time
NIBP measurements become outline fonts after a preset time. This avoids older NIBP
values being misinterpreted as current measurements. To set the timeout period, follow