P1 Patient Monitor User’s Manual
12-8
1. Select the NIBP parameter area to enter the【NIBP】menu.
2. Select【Alarm】submenu.
3. If the alarm setting is protected by password, enter the password. For detail,
please refer to 7.7.2 Changing Alarm Setup Protection Mode.
4. Set alarms as needed.
12.9.2. Setting the Initial Cuff Inflation Pressure
You can manually set the initial inflation pressure of the cuff, follow this procedure:
1. Select the NIBP parameter area to enter the【NIBP】menu.
2. Select【Setup】submenu.
3. Set【Initial Pressure】:
Select the appropriate cuff pressure value as needed.
12.9.3. Setting the NIBP Interval
For Auto NIBP measurement, you need to set the interval between two NIBP
measurements. To set the NIBP interval, follow this procedure:
1. Select the NIBP parameter area or waveform area to enter the【NIBP】menu.
2. Select【Setup】submenu.
3. Set【Interval】.
12.9.4. Selecting NIBP Start Mode
Start mode defines how NIBP auto mode works. To set the 【Start Mode】, follow
this procedure:
1. Select the NIBP parameter area a to enter the【NIBP】menu
2. Select【Setup】submenu.
3. Set【Start Mode】.
【Clock】: After the first measurement, the monitor automatically
synchronizes NIBP automatic measurements with the real time clock.
For example, if 【Interval】 is set to 【30min】, and you start NIBP auto
measurement at 10:03, the next measurement will be taken at 10:30, and
then at 11:00, 11:30, and so on.