P1 Patient Monitor User’s Manual
36
Appendix C Default Settings
This chapter lists some important factory default settings for monitors. The user
cannot change the factory default settings, but the monitor can be restored to the factory
default settings when necessary.
C.1 ECG, Arrhythmia, ST and QT Default Settings
C.1.1 ECGDefault Settings
Item Default Setting
HR Alarm switch ON
Alarm high limit Adult: 120 bpm
Pediatric: 160 bpm
Neonatal: 200 bpm
Alarm low limit Adult: 50 bpm
Pediatric: 75 bpm
Neonatal: 100 bpm
Alarm priority Med
Alarm print Off
Alarm source HR
ECG 1 II
ECG2 (5-lead, 6-lead, 12-lead) I
Gain ×1
Wave Speed 25 mm/s
Filter Mode Monitor
Notch Filter On
Lead type 3-lead
QRS volume 3
Paced Adult: Unspecified
Pediatric/Neonatal: Turn off
Pacer Reject Off