P1 Patient Monitor User’s Manual
9-5
9.5.3. Selecting Respiration Lead
You can set up respiration lead to get the best respiratory waveform. The steps to set
up breathing leads are as follows:
1. Select the RESP parameter area or waveform area to enter the RESPmenu.
2. SelectSetupsubmenu.
3. Set RESP Lead.
If the respiratory waveform is still poor after adjusting the
respiration lead or the respiration rate measurement is suspected to be
inaccurate, you can adjust the electrode position.
9.5.4. Setting RESP Waveform Gain
You can adjust the RESP waveform gain to better view the waveform amplitude. The
steps to set the RESP waveform gain are as follows:
1. Select the RESP parameter area or waveform area to enter the RESPmenu.
2. SelectSetupsubmenu.
3. SetGain.
9.5.5. Setting the RESP Waveform Speed
To set the RESP waveform speed, follow this procedure:
1. Select the RESP parameter area or waveform area to enter the RESPmenu.
2. SelectSetupsubmenu.
3. SetWave Speed.
9.5.6. Setting the Auto Detection Switch
To set the auto detection switch, follow this procedure:
1. Select the RESP parameter area or waveform area to enter the RESPmenu.
2. SelectSetupsubmenu.
3. Switch on or offAuto Threshold Detection.
If Auto Threshold Detection is switched on, the monitor
automatically adjusts the RESP waveform detection level, or threshold.
If Auto Threshold Detectionis switched off, you have to manually