P1 Patient Monitor User’s Manual
38
C.1.3 ST Default Settings
Item Default Setting
ST-I, ST-II, ST-III,
ST-aVR, ST-aVL,
ST-aVF, ST-V1,
ST-V2, ST-V3,
ST-V4, ST-V5,
ST-V6, ST-Va,
ST-Vb
Alarm switch ON
Alarm high limit 0.2 mV
Alarm low limit -0.2 mV
Alarm priority MED
Alarm print OFF
ST Analysis OFF
ST Mark OFF
Auto Adjust OFF
ST Point J + 60 ms
ISO -80 ms
J 48 ms
C.1.4 QT Default Settings
Item Default Setting
QTc Alarm switch OFF
Alarm high limit Adult: 500
Pediatric: 480
Neonate: 460
Alarm priority MED
Alarm print OFF
QTc Alarm switch OFF
Alarm high limit 60
Alarm priority MED
Alarm print OFF
QT Analysis OFF