P1 Patient Monitor User’s Manual
59
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
Out of Range pressure.
CO
2
Span Calibration
Command Failed
Low Not clearable
Contact the after-sales personnel
to check the maintenance.
CO
2
Span Calibration
in Progress…
Prompt / Disappear after success.
CO
2
Replace Adapter Low Not clearable Check adapter.
CO
2
No Adapter Low Completely clear Check adapter.
D.2.9 System Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
Battery Low High Not clearable
Please connect AC power supply for
power supply and charge the battery.
Battery Fault Prompt / Please replace the battery.
CMS Disconnected Med
Completely
clear
The monitor is disconnected from the
CMS. Please check the network
connection.
Disk Full Med Not clearable
The storage space of the monitor is full.
Please clear the patient related data in
time.
Disk Will Be Full Low
Sound and light
can be cleared
The storage space of the monitor is
almost full. Please clear the patient
related data.