P1 Patient Monitor User’s Manual
58
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
CO
2
Check Sampling
Line
Low Not clearable
The CO
2
sampling tube is
blocked or damaged; The
sampling tube is kinked or
compacted; The exhaust pipe is
blocked. Check the sampling
tube.
CO
2
Check Adapter Low Not clearable Reinstall the airway adapter.
CO
2
Zero In
Progress…
Prompt / The CO
2
module is being zeroed.
CO
2
Sensor Warm up Prompt / The CO
2
module is warming up.
CO
2
Self-Test Prompt / Module initialization
CO
2
Sensor Faulty, E* Low /
Hardware or software errors,
contact after-sales personnel to
check maintenance.
CO
2
Self-Test Error Low Not clearable
Hardware errors, Replace sensor,
if the problem cannot be solved,
please contact the after-sales
personnel to check the
maintenance.
CO
2
Motor Speed
Error
Low Not clearable
Check whether the sampling tube
is blocked.
CO
2
Factory
Calibration lost
Low Not clearable
Contact the after-sales personnel
to check the maintenance.
CO
2
Sampling Line
Clogged
Low Not clearable Check sampling line.
CO
2
No Sample Line Low Completely clear Check sampling line.
CO
2
Internal Temp.Out
of Range
Low Not clearable
Hardware error, and contact the
after-sales personnel to check the
maintenance.
CO
2
Ambient Pressure Low Not clearable Recalibrate atmospheric