P1 Patient Monitor User’s Manual
57
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
sensor is not connected or the
pressure is out of range or the
pressure is unstable.
XX Catheter Off HIGH Not clearable
The catheter is pulled out from the
patient. Please check the
connection.
Zero Required Prompt / /
Zero Succeed Prompt /
The IBP module is zeroing
successful
Note: XX represents IBP labels, such as PA, CVP, FAP, P1, etc.
D.2.8 CO
2
Technical Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
CO
2
Sensor Off Med Completely clear
The CO
2
sensor is detached from
the patient or monitor.
CO
2
Out of Range Low Not clearable
The measured data of CO
2
module is out of range and needs
zero.
CO
2
Zero Required Low Not clearable The sensor needs zero.
CO
2
Sensor Over
Temp
Low Not clearable Check sensor.
CO
2
Compensation
Not Set
Low Not clearable
The CO
2
sensor was not
initialized. Set compensation and
initialize.
CO
2
Sleep Mode Prompt /
The CO
2
sensor is in sleep mode.
Please select the measurement
mode, CO
2
can enter the working
state.