P1 Patient Monitor User’s Manual
56
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
Or irregular pulse rate, such as
arrhythmia.
NIBP Signal
Saturated
Low
Completely
clear
Great movement.
NIBP Weak
Signal
Low
Completely
clear
The cuff is too loose or the patient’s
pulse is too weak. Please check the
patient’s condition or whether the cuff is
placed correctly.
NIBP Out of
Range
Low
Completely
clear
The measurement range exceeds the
specified upper limit.
NIBP Time Out Low
Completely
clear
Measurement time exceeds 120s
(adult/pediatric) or 90s (neonatal).
Please check the patient’s condition and
the connection of accessories or replace
the cuff, and conduct the measurement
again.
NIBP Cycle Abort Low
Completely
clear
Three consecutive measurement failures
occurred during periodic measurement.
Please check whether the patient’s
condition or cuff placement is correct.
NIBP Zero Failed Prompt /
At zero, the pressure is beyond the zero
range or the pressure is unstable.
D.2.7 IBP Technical Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
IBP Self-test Error Med Not clearable
The board is failure. Please
contact the factory for repair.
XX Sensor Off Med Completely clear The XX cable is off the monitor.
XX Zero Failed Med Not clearable When the XX sensor is zeroed, the