P1 Patient Monitor User’s Manual
17-4
17.2.6. HL7 Settings
The real-time data, waveforms, and alarms of the monitor can be transmitted to
the hospital’s monitoring system through the HL7 protocol. The operating steps are as
follows:
1. Select 【 Main Menu 】 quick keyfrom 【 System 】 column to select
【Maintenance】input maintenance passwordEnter.
2. Select 【Network Setup】 submenu 【HL7】 submenu.
3. Select 【Parameter】, 【Waveform】and 【Alarm】sending function as
required.
From 【Physiological data】 column to select monitor as 【Server】
or 【Client】. If select the monitor as 【Client】, set the 【IP】 and
【Port】 for the server receiving the real-time data and waveform. And
can set 【Interval】of data.
From 【Alarm Data】 column to select monitor as 【Server】 or
【Client】. If select the monitor as 【Client】
, set the 【IP】 and 【Port】
for the server receiving the real-time data and waveform.
17.2.7. Connecting eGateway
The monitor can connect the eGateway server through wired and wireless
networks to realize the interaction between the external devices and the monitor. The
monitor has the following functions when connected to eGateway:
Send parameters, waveforms, and events of this monitor to eGateway.
Send data on external devices connected to the monitor to eGateway,
including parameters, alarms, etc.
Clock can be synchronized between the monitor and the eGateway.
17.2.8. Using the ADT Gateway
The ADT (admit-discharge-transfer) gateway is normally deployed in the
eGateway. You can obtain patient information from the hospital ADT server through
the ADT gateway.