EasyManua.ls Logo

Biolight Meditech P1 - Page 128

Default Icon
303 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
P1 Patient Monitor User’s Manual
10-4
SpO
2
and also partially indicates the signal quality.
Above 1 is optimal;
Between 0.3 and 1 is acceptable;
Below 0.3 indicates low perfusion. When 0.3PI<1, the PI values will be
displayed in yellow background; PI<0.3, PI values will be displayed in red
background, and SpO
2
parameter values are displayed as hollow words.
Reposition the SpO
2
sensor or find a better site. If low perfusion persists,
choose another method to measure oxygen saturation if possible.
(5) Pulse rate: the number of pulses detected per minute (from the pleth waveform)
(6) ∆Sp
:
󰇛
∆Sp

|

Sp
|󰇜
For Masimo module
The Masimo module is intended to monitor the SpO
2
, SpMet, PVI, SpHb, SpOC, PI,
SpCO of patients. The following figure shows the SpO
2
display screen, the display on
your monitor may be looked slightly different.
1. SpO
2
value;
2. SpO
2
unit;
3. SpMet (Methemoglobin Saturation) value and unit;
4. PVI (Pleth Variability Index) value and unit;
5. SpHb (Total Hemoglobin) value and unit;
6. SpOC (Oxygen Content) value and unit;
7. PI (Perfusion Index) value and unit;
8. SpCO (Carboxyhemoglobin Saturation) value and unit.
678
2
3 4 5
1

Table of Contents