P1 Patient Monitor User’s Manual
14-11
14.6.4. Setting the CO
2
Waveform
To set the CO
2
waveform, follow this procedure:
1. Select the CO
2
parameter area or waveform area to enter theCO
2
menu.
2. SelectSetupsubmenu.
3. Set CO
2
waveformWave Mode, Wave SpeedorScale.
14.6.5. Setting CO
2
Corrections
Temperature, water vapor in the patient’s breath, barometric pressure, and the
proportions of O
2
, N
2
O and Helium in the mixture all influence CO
2
absorption.
For mainstream and sidestream/microflow CO
2
module, you can set the CO
2
correction by following the steps below:
1. Select the CO
2
parameter area or waveform area to enter theCO
2
menu.
2. SelectSetupsubmenu.
3. Please set the following contents according to the actual situation before
correction:
——Gas Temperature: Set the temperature of gas.
——Barometric Pressure: Set the atmospheric pressure.
The system default barometric pressure is 760mmHg, you can selectMain Menu
quick keyfromSystemcolumn to selectMaintenancein
Other
submenu to
enter
the barometric pressure value of the environment.
——Zero Gas Type: Select the gas type of zeroing, the options areZero On
Room AirorZero On N
2
.
——O
2
Compensation: Select the concentration of oxygen. It can be set to a
value between 0% and 100%. The default value is 16%.
——Anesthetic Gas: Select the concentration of anesthetic agent. It can be set to
a value between 0.0% and 20.0%. The default value is 0.0%.
——Balance Gas:
Select the type of balance gas, the options are Room Air,
N
2
OorHelium.
When the most proportions of the mixture is air, selectRoom AirWhen the most
proportions of the mixture is N
2
O, selectN
2
O. When the most proportions of the