P1 Patient Monitor User’s Manual
15-5
Delete parameters: select desired parameter from the 【Selected】
column on the right and then select 【Delete】.
Move the position of parameters: select desired parameters from the
【Selected】 column on the right and select 【Move Up】, 【Move
Down】, 【Move To Top】 or 【Move To Bottom】.
Selecting 【Default Config】 will resume the trend group setting to factory defaults.
CAUTION:
When 【Trend Group】 is set to 【All】 or 【Standard】, you cannot edit the
trend group.
ECG parameter and waveform are always displayed in the first row on the trend
page. It cannot be deleted or moved.
15.2.5.4. Changing the Resolution of Trend Data
The resolution of tabular trends defines the interval of displaying trend data. High
resolution is especially suited for neonatal monitoring, where the clinical situation may
change very quickly. In adult monitoring, where the patient’s status typically changes
more gradually, a low resolution may be more informative.
To change resolution, follow this procedure:
1. Enter the tabular trends review page.
2. Select 【…】 to enter setup menu.
3. Select 【Sample Rate】.
【5s or 30s】:The trend of parameters in the last 6 hours was observed
at intervals of 5 seconds or 30 seconds.
【1 min、5 min、10min、15min、30min、1h、2h、3 h】:According
to the selected time interval, observe the parameter trend of the last 180
hours.
【NIBP】: The tabular trends show the values of each parameter at the
measurement time of NIBP parameters.