P1 Patient Monitor User’s Manual
16-3
dose.
INF Rate: The titration table is listed in the sequence of increased
infusion rate.
16.3.3. Drug calculation Formula
Description Unit Formula
Drug Amount
g series: mcg, mg, g;
unit series: unit, kU, MU;
mEq series: mEq;
Dose Amount = Liquid Volume
×Drug concentration
Liquid
Volume
ml Manual input required
Drug
concentration
mcg/ml, mg/ml, g/ml,
Unit/ml, kU/ml, MU/ml,
mEq/ml
Drug concentration = Dose
Amount / Liquid Volume
Drop Size GTT/ml Manual input required
Dose/hr
g seriesmcg, mg, g;
unit series: unit, kU, MU;
mEq series: mEq;
Dose/h=Dose/min×60
Dose/min
g seriesmcg, mg, g;
unit series: unit, kU, MU;
mEq series: mEq;
DoseMin DoseMin
Dose/kg/hr
(weight based)
g seriesmcg, mg, g;
unit series: unit, kU, MU;
mEq series: mEq;
Dose/h= Dose/h/weight
Dose/kg/min
(weight based)
g seriesmcg, mg, g;
unit series: unit, kU, MU;
mEq series: mEq;
Dose/min= Dose/min /weight
INF rate ml/h
Infusion rate = Dose/h/drug
concentration