P1 Patient Monitor User’s Manual
34
b
Over the frequency range 0.15 MHz to 80 MHz, field strengths should be less than 3 V/m.
Table 4
Recommended separation distances between portable and mobile RF communications
equipment and the monitor
The monitor is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the monitor can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the monitor.
Rated
maximum
output power of
transmitter (w)
Separation distance according to frequency of transmitter
150kHz80MHz
d =
P2.1
80MHz800MHz
d =
P2.1
80 MHz2.7 GHz
d =
P3.2
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.20 1.20 2.30
10 3.80 3.80 7.30
100 12.00 12.00 23.00
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.