P1 Patient Monitor User’s Manual
2-2
entanglement or suffocated, the cables entangled, or subject to electrical
interference.
To avoid danger or environment pollution, the packaging materials must be
handled in accordance with local regulations or the hospital's waste disposal
system. Packaging materials must be placed out of reach of children.
Ensure that the equipment is supplied with continuous electric power during
work. Sudden power failure may cause data loss.
The user should periodically check and move the sensor on the skin to avoid
adverse skin or tissue effects.
CAUTION
To ensure patient safety, use only parts and accessories specified in this
manual.
Magnetic and electrical fields are capable of interfering with the proper
performance of the monitor. For this reason make sure that all external devices
operated in the vicinity of the monitor comply with the relevant EMC
requirements. Mobile phone, X-ray equipment or MRI devices are a possible
source of interference as they may emit higher levels of electromagnetic
radiation.
To avoid contamination or infection of personnel, environment or other
equipments, the equipment and its accessories that meet the service life must be
disposed in accordance with relevant local regulations or hospital systems.
The lifetime of the patient monitor is 5 years.
Before connecting the equipment to the power line, check that the voltage and
frequency ratings of the power line are the same as those indicated on the
equipment’s label or in this manual.
Always install or carry the equipment properly to avoid damage caused by
drop, impact, strong vibration or other mechanical force.
The monitor must be wiped dry immediately after exposure to rain or splashes.
Please do not mix different types and brands of electrodes. Mixing the
electrodes may result in a large baseline drift or a long baseline recovery time
after defibrillation. It is forbidden to use dissimilar metal electrodes, which