P1 Patient Monitor User’s Manual
54
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
connection between the SpO
2
sensor and the SpO
2
main cable
falls off. Confirm that SpO
2
main
cable and sensor are connected
normally. If the alarm still cannot
be eliminated, replace the sensor.
SpO
2
Low
Confidence
Low Not clearable PI<0.3% or signal quality <60.
SpO
2
Update
Timeout
Low Not clearable 25s SpO
2
measurement data not
updated.
SpO
2
Motion
Interference
Low Not clearable Patients move too much, affecting
measurement.
SpO
2
Searching Pulse Prompt
/ SpO
2
module is searching for
pulse.
SpO
2
Sensor
Incompatible
Med Not clearable
Non-factory SpO
2
sensor are used.
Please replace the original sensor.
SpO
2
Sensor Has
Expired
Med Not clearable
SpO
2
sensor has expired. Please
replace the sensor.
SpO
2
Sensor is A
out
to Expire
Prompt /
SpO
2
sensor is about to expire.
Please replace the sensor in time.
D.2.5 TEMP Technical Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
TEMP Self-test
Error
Med Not clearable
Board failure. Please contact the
manufacturer for repair.
〈TEMP label〉
Sensor Off
Med
Completely
clear
Check the connection of the sensor and
reconnect the sensor.