P1 Patient Monitor User’s Manual
7-3
7.3.3. Alarm Indicators
When an alarm occurs, the patient monitor will indicate it to the user through visual
or audible alarm indications:
Alarm signal
High priority
alarm
Medium priority
alarm
Low priority
alarm
Prompt Note
AlarmLamp
The lamp quickly
flashes red with
1.4Hz ~2.8Hz,
Duty cycle
20%-60%.
The lamp slowly
flashes yellow
with 0.4Hz~
0.8Hz, Duty
cycle 20-60%.
The lamp turns
cyan without
flashing, Duty
cycle 100%.
/ /
Alarm Tone
Mode
ISO
DO-DO-DO
--
DO-
DO
-----
DO-DO-DO
--
DO-DO
-----
DO-DO-DO - DO- / /
Alarm Information
White words,
Red background
Black words,
Yellow
background
Black words,
Cyan
background
White
words
Displayed in the
top information
area, click on
the alarm
information to
view the alarm
information list.
Alarm level
symbol
*** ** * /
Displayed in
front of alarm
information.
Parameteralarm
Red background,
flashes
Yellow
background,
flashes
Cyan
background,
flashes
/ /
WARNING:
When multiple alarms of different priorities occur simultaneously, the alarm
lamp and alarm tone are prompted according to the highest level of all current
alarms.