P1 Patient Monitor User’s Manual
8-13
NOTE
Only the Filter Modeis set to Diagnose Mode or ST can you switch
on or offNotch Filter, other mode is enabled by default.
8.5.5.5. Setting Notch Filter Frequency
According to the mains frequency of your country, you can set the frequency of the
notch to50Hz or60Hz. If you need to change the Notch Frequency, please
contact the manufacturer maintenance personnel.
8.5.6. Setting Smart Lead Switch
This monitor provides the function of switching main lead automatically. When
switch on Smart Lead(Smart lead auto switchover), the current smart leads are
automatically identified by the algorithm, and the host automatically switches the smart
leads according to the identification of the algorithm.
Steps of switching off smart lead
function are as follows:
1. Select ECG parameter area or waveform area to enter theECGmenu.
2. SelectSetupsubmenu.
3. Switch offSmart Lead.
8.5.7. Setting the Priority of the ECG Lead Off Alarm
The steps to set the alarm level for ECG lead off alarms are as follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuOthersubmenu.
3. SetECG
Lead Off Alarm Level.
8.5.8. Adjusting the QRS Volume
The QRS volume is determined byAlarm Sourcein the ECG or PR alarm setting
menu. Which parameter (HR or PR) is set toAlarm Sourceand the QRS volume is