BIOPAC Hardware | ELSTM Series | Page 1 - 1 Updated: 5.14.2013
ELSTM1 UNSHIELDED STIMULATING BAR ELECTRODE
The ELSTM1 Unshielded Stimulation Bar Electrode for Research Systems has a BNC connector that will
interface with either the STM200 or the STMISOLA.
ELSTM1 SPECIFICATIONS
Bar Length: 4 cm
Spacing between contacts: 3 cm
Electrode Contact Diameter: 1 cm
Cable Length: 2 m
ELSTM2 UNSHIELDED NEE
DLE ELECTRODES
Reco
mmended for use when applying a stimulus to animal subjects and tissue preparations. The dual stainless
steel needles are Teflon coated. The coating prevents the needle from making contact with the subject except at
the very tip of the needle, which is exposed. For applications that require better contact between the electrode and
the subject to record a good
signal, abrade the needle to remove the Teflon coating. Needle electrodes are shipped
non-sterile, so pre-sterilization is required.
ELSTM2 SPECIFICATIONS
Needle Length: 2.5 cm
Needle Diameter: 0.3 mm
Cable length: 2.5 m
Connector type: BNC
Interface: BSLSTM Stimulator or SS58L for MP35 or OUT3 for MP36