BIOPAC Hardware | SS56L | Page 1 - 1 Updated: 2.12.2013
SS56L HAND CLENCH FORCE BULB
SS56L measures proportionality of bulb pressure to clench
force in “kgf/m^2” units (a pressure unit). This measure is
accurate for the relative measures recorded in BSL Lesson 2
Electromyography (EMG) II. SS56L is recognized by current
release BSL Lessons.
SPECIFICATIONS
Pressure Range:
BSL: 0 to 10,546 Kgf/m^2
AcqKnowledge: 0 to 1.0546 Kg-f/cm^2
0 to 15 psi
Accuracy: ±3%
Output:
BSL: 0.58 mV/100 Kgf/m^2
AcqKnowledge: 0.58 mV/0.01 Kg-f/cm^2
4.1 mV/psi
Bulb Diameter: 5.8 cm
Bulb Length: 11.1 cm
Tubing Length: 3 meters
Weight: 108 g
Optional BSL PRO Presets:
Clench Force - kpa (SS56L) - input value 20.48 mv scales to 34.47 kpa
Clench Force - psi (SS56L) - input value 61.44 mv scales to 15 psi