EasyManua.ls Logo

BK Precision 2194 - Page 36

BK Precision 2194
139 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Sample Control 36
Disabled:
When XY is disabled, YT mode is set. In YT mode the display is set to a volt versus time graph, and signal events
occurring before the trigger are plotted to the left of the trigger point and signal events after the trigger plotted to the
right of the trigger point.
Enabled:
XY mode changes the display to a volt versus volt graph. Channelโ€™s 1 amplitude is plotted on the x-axis and channelโ€™s
2 amplitude is plotted on the Y-axis.
The phase deviation between two signals with the same frequency can be measured via the Lissajous method. The ๎™ฝgure
below shows the measurement schematic diagram of the phase deviation.
Figure 5.10 Phase Deviation
According to ๐‘ ๐‘–๐‘›(๐œƒ) = (
๐ด
๐ต
) ๐‘œ๐‘Ÿ (
๐ถ
๐ท
) (where is the phase deviation angle between the two channels and the de๎™ฝnitions of A,
B, C and D are as shown in ๎™ฝgure 5.10), the phase deviation angle is obtained using ๐œƒ =
ยฑ
๐‘Ž๐‘Ÿ๐‘๐‘ ๐‘–๐‘› (
๐ด
๐ต
) ๐‘œ๐‘Ÿ
ยฑ
๐‘Ž๐‘Ÿ๐‘๐‘ ๐‘–๐‘› (
๐ถ
๐ท
).
If the principal axis of the ellipse is within quadrant I and III, the phase deviation angle obtained should be within quadrant
I and IV, namely within (0 to ฯ€/2) or (3ฯ€ /2 to 2ฯ€).
If the principal axis of the ellipse is within quadrant II and IV, the phase deviation angle obtained should be within
quadrant II and III, namely within (ฯ€ /2 to ฯ€) or (ฯ€ to 3ฯ€/2).

Table of Contents

Related product manuals