Responding to Error Messages
154
5. Troubleshooting
If the Error Occurs Again
• The collet is deformed�
Procedure
A
Replace the collet.
"
B
Perform automatic correction.
"
"101D-010*" The % milling bur cannot be released. It might be broken from the root.
* The milling bur stocker number (1 to 7) is displayed for %�
Level 2
Situation/Error Cause
• The milling bur is broken�
• The milling bur has exceeded its service life�
"
Procedure
A
Follow the instructions in the VPanel window to open the collet and remove the milling bur.
B
Replace the milling bur with a new one.
C
Perform collet maintenance.
"
D
Perform automatic correction.
"
Situation/Error Cause 2
• The milling conditions are too strict�
Procedure
A
Hold down the operation button on the built-in panel.
This will clear the error.
B
Review the CAM settings and the shape specied in the CAD data.