Responding to Error Messages
166
5. Troubleshooting
"103B-0000" The automatic correction is not yet nished.
Level 2
Situation/Error Cause
Automatic correction has not been performed�
Automatic correction was cancelled before it could nish�
Automatic correction was not performed after updating the rmware to a version that required automatic correction to be
performed again�
The versions of VPanel and the machine's rmware do not match�
Procedure
A
Hold down the operation button on the built-in panel.
This will clear the error.
B
Download the latest versions of VPanel and the machine's rmware, and then install these versions.
If you have already upgraded the version, proceed to the next step.
" DGSHAPE Corporation website (http://www.dgshape.com/)
C
Perform automatic correction.
" 
"103D-0000" Milling data error. The milling bur cannot reach the milling position.
Level 2
Situation/Error Cause
The milling bur is too short to reach the milling position�
In the milling data that has been sent, the movement in the negative direction along the Z axis is outside the movement
range�
Procedure
A
Hold down the operation button on the built-in panel.
This will clear the error.
B
Check the milling data.
C
Review the milling bur length and the position of the milling bur holder.
" 