Appendix
3 = SPAN
86
Primary Raw Conc.
(For S40, before NOx background and gain)
87
Secondary Raw Conc.
Only for multigas instruments
(For S40, before NOx background and gain)
88
S40 Backgnd Conc.
NOx Background Concentration
(For S40, before gain)
89
Cal. Pressure
Calibration Pressure
90
Conv. Efficiency
Converter Efficiency
91
Multidrop Baud Rate
0 = 1200 bps
1 = 2400 bps
2 = 4800 bps
3 = 9600 bps
4 = 14400 bps
5 = 19200 bps
6 = 38400 bps
92
Anlg Range AO 0
Maximum range value for analog output
93
Anlg Range AO 1
94
Anlg Range AO 2
95
Output Type AO 0
Output Type
1 = Voltage
0 = Current
96
Output Type AO 1
97
Output Type AO 2
98
Anlg Ofst/Rng AO 0
Voltage Offset /Current Range
0 = 0% or 0 - 20mA
1 = 5% or 2 - 20mA
2 = 10% or 4 - 20mA
99
Anlg Ofst/Rng AO 1
100
Anlg Ofst/Rng AO 2
101
F/Scale Volt AO 0
5.0 Volt Calibration value
102
F/Scale Volt AO 1
103
F/Scale Volt AO 2
104
Z Adj Volt AO 0
0.5 Volt Calibration value
105
Z Adj Volt AO 1
106
Z Adj Volt AO 2
107
-Analog Supply
Negative analog supply
108
Digital Outputs
A single byte expressing the most recent state of the digital
outputs
109
Digital Inputs
A single byte expressing the most recent state of the digital
inputs
110
Instrument State
0 = Sample Fill
1 = Sample Measure
2 = Sample Fill Aux

Table of Contents