Pressure Temperature Flow Compensation Factor for third gas
in multi-gas instruments.
The current dilution ratio (default is 1.0)
State of the status light:
0 = Green
1 = Amber
2 = OFF (normally impossible)
3 = Red
0 = EC9800
1 = Bayern-Hessen
2 = Advanced
3 = Modbus
An offset applied to Gas 4
Ozone generator control, DAC controlled.
DAC: 0..64535
The data logging period, in seconds (1.. 86400)
Ozone generator coefficients
Note that Coeff D is parameter 127
Only valid with an S44 variant set to multiple gains. Shows the
third gas gain (Nx). Default is 1.0.
Only applies to S40 variants.
0 = new PTF calculation
1 = old PTF calculation inherited from ML
This value was added to Build 10800 and defaults to 1.
Only applies to S30 variants, and only on Rev P or later boards.
The current humidity inside the instrument, in percentage.
Only applies to Rev P or later boards.
Optical signal strength.