Acclarix LX9 Series Diagnostic Ultrasound System User Manual
regulation.
The system does not need calibration as part of routine maintenance.
The format of U disk should be FAT32.
Do NOT place the device on slopes. It may suddenly slide, resulting in injury and/or equipment
damage.
Do NOT stand/sit on or bend over the device. It may move and make you lose your balance
and tumble.
To ensure safety, two persons are required to move the device across slopes.
Transducer Cautions
Do not use disinfection agents beyond their expiration date.
Do not use sterile sheaths beyond their expiration date.
Inspect the transducer connector, cable, and head periodically. Do not use if there is evidence
of excessive wear or damage.
Do not operate the transducer to temperatures in excess of 40°C or store the transducer in
temperatures in excess of 55°C.
Do not kink or pull on the transducer cable.
Broken or bent connector pins can cause image artifacts. Do not use a transducer with broken
or bent pins.
Network Security Cautions
Keep your ultrasound system safe to protect the patient information and data from being
modified, damaged or disclosed caused by unauthorized disassembly.
Always ensure the privacy of patient information and data displayed/stored in the ultrasound
system or exported to external storage devices.
The software upgrade can only be performed by EDAN-qualified service professionals with
upgrade files of known provenance. Confirm that the system boots to imaging after an
upgrade.
Make sure the ultrasound system is used under secure network environment, and all the
approved devices connecting with the ultrasound system are physically secure.
Anti-virus measures such as USB device virus scanning should be carried out prior to using
the USB flash drive.
Do not connect an USB device with unknown provenance to the ultrasound system.
When the ultrasound system is returned for maintenance, disposed of, or removed from the
medical institution for other reasons, ensure all patient data are removed from the ultrasound
system.
Always hide your password when entering it on the system, protecting it from being disclosed.
Federal Communications Commission (FCC) Statement:
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions:
This device may not cause harmful interference, and
This device must accept any interference received, including interference that may cause
undesired operation.
This equipment has been tested and found to comply with the limits for a Class B digital device,