Acclarix LX9 Series Diagnostic Ultrasound System User Manual
To use gel warmer:
After installing the gel warmer, put a bottle of ultrasound gel inside the gel warmer, and press the
power switch on the gel warmer to start heating.
The gel warmer can heat the gel to about 36±2℃. It will take about 0.5h to heating a bottle of 250ml
ultrasound gel from 18℃ to 28℃.
The gel warmer supports over-temperature protection and it can work continuously over 12 hours.
1. When the over-temperature system of the gel warmer fails, the ultrasound system will display a
prompt message "Gel heater is abnormal, there is overheating risk. Please stop using and
contact the supplier for maintenance." In this case, you should switch off the gel warmer
immediately and contact the service for maintenance.
2. Do not power on the gel warmer when no ultrasound gel is put inside, otherwise the gel warmer
will be abnormal.
3. If you feel the gel is obviously too hot, please stop using it and contact the serviceman.
4. The gel warmer does not support "hot plug" feature. Please install the gel warmer before starting
the ultrasound system. Otherwise the gel warmer may be unrecognized or works abnormally.
Note:
Do not spray any liquid on or into the gel warmer in use.