Acclarix LX9 Series Diagnostic Ultrasound System User Manual
To clean gel warmer
1. Switch off the system and disconnect the gel warmer.
2. Rotate to remove the gel warmer bottom cover.
3. Use a sterile cloth dampened with cleaning agent to gently wipe the entire exterior
surface(including the bottom cover) and inner surface of the warmer thoroughly until no visible
contaminants remain.
4. After cleaning, wipe off the cleaning agent with a new sterile cloth dampened with tap water until
no visible cleaning agent remains.
5. Leave the gel warmer air dry.
The validated cleaning agents for cleaning the gel warmer are:
Ethanol (75%)
Isopropanol (70%)
Note:
It is recommended to use the validated cleaning agents to clean the gel warmer. Do not use cloth with
chemicals or use sharp-edged material (like steel wool) to clean the gel warmer.