Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block
Endress+Hauser 127
Sys. - Positive Zero
Return
AUTO - OOS Use this parameter to interrupt evaluation of measured variables. This is
necessary when a piping system is being cleaned, for example. This setting
acts on all parameter and outputs of the measuring device.
Options:
OFF
ON (signal output is set to the ZERO FLOW value, temperature and density
are output normally)
Factory setting:
OFF
! Note!
The active positive zero return is passed on to subsequent function blocks
via the status condition (UNCERTAIN) of the process variable.
Interruption of the processing of measured values can, also using cyclic
scheduled data transmission, be controlled or initiated via the Discrete
Output function block.
Sensor Data -
K-Factor
read only This parameter shows the current calibration factor for the sensor.
Factory setting:
Depends on nominal diameter and calibration.
! Note!
If the service code is used to call this parameter, this value can be edited.
Sensor Data -
Zeropoint
read only This parameter shows the current zero-point correction value for the sensor.
Display:
Max. 5-digit number: –99999 to +99999
Factory setting:
Depends on calibration.
Sensor Data -
Nominal Diameter
read only This function shows the nominal diameter for the sensor.
Factory setting:
Depends on nominal diameter
! Note!
If the service code is used to call this parameter, this value can be edited.
Sensor Data -
Temp. Coeff.
KM
read only This function shows the temperature coefficient KM.
Sensor Data -
Temp. Coeff.
KM2
read only This function shows the temperature coefficient KM2.
Sensor Data -
Temp. Coeff. KT
read only This function shows the temperature coefficient KT.
Sensor Data -
Cal. Coeff. KD1
read only This function shows the calibration coefficient KD1.
Sensor Data -
Cal. Coeff. KD2
read only This function shows the calibration coefficient KD2.
"Flow" Transducer Block / base index 1400:
Parameter
Write access
with operating
mode
(MODE_BLK)
Description