Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block
Endress+Hauser 179
Tube Damp.Fluct. -
Ref.Value Factory
read only The reference value for the fluctuation of the tube damping appears on the
display.
Display:
5-digit floating-point number
Tube Damp.Fluct. -
Ref.Value User
read only The reference value for the fluctuation of the tube damping appears on the
display.
Display:
5-digit floating-point number
Tube Damp.Fluct. -
Actual Value
read only The measured fluctuation of the tube damping appears on the display.
Display:
5-digit floating-point number
Tube Damp.Fluct. -
Min.Value
read only The lowest value of the tube damping since the saved values were last reset
appears on the display.
Display:
5-digit floating-point number
Tube Damp.Fluct. -
Max.Value
read only The highest value of the tube damping fluctuation since the saved values
were last reset appears on the display.
Display:
5-digit floating-point number
Tube Damp.Fluct. -
History 1…10
read only The last ten values of the tube damping fluctuation since the saved values
were last reset appear on the display.
Display:
5-digit floating-point number
Tube Damp.Fluct. -
Deviation
read only This function displays the deviation between the measured fluctuation in
tube damping and the reference values selected in the parameter "Adv. -
Select Ref.Condition" (FACTORY or USER)Page 170.
Display:
5-digit floating-point number
Tube Damp.Fluct. -
Warning Level
AUTO - OOS
! Note!
This option is not active unless ON was selected in the
parameter "Adv. - Warning Mode".
Use this parameter to specify a limit value for tube damping fluctuation. A
notice message is generated if the limit value is exceeded.
User input:
0 to 99999
Factory setting:
1000
Amp. Device Type read only This parameter is only used for service purposes.
"Advanced Diagnostics" Transducer Block/base index 2200
Parameter
Write access
with operating
mode
(MODE_BLK)
Description