3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus
170 Endress+Hauser
Adv. -
Ref.Cond.User
AUTO - OOS Use this parameter to start determining the user reference status. The
following values are determined:
•MASS FLOW
•DENSITY
•REFERENCE DENSITY
TEMPERATURE
TUBE DAMPING
ELECTRODYNAMIC SENSORS
OPERATING FREQUENCY FLUCTUATION
TUBE DAMPING FLUCTUATION
Options:
CANCEL
START
Factory setting:
CANCEL
Adv. - Select
Ref.Condition
AUTO - OOS Use this parameter to select the reference status which should be used to
compare the advanced diagnostics parameters (see parameter
"Adv. - Acquisition Mode"Page 171).
Options:
FACTORY
USER
Factory setting:
FACTORY
Adv. -
Warning Mode
AUTO - OOS Use this parameter to determine whether a warning should be generated
when there is a deviation between the reference status (FACTORY or USER,
see parameter "Adv. - Select Ref.Condition") and the current measuring
values.
The values of the following parameters are compared to the reference status :
Mass Flow - Actual Value ( Page 172)
Density - Actual Value (Page 173)
Ref.Density - Actual Value (Page 174)
Temperature - Actual Value (Page 175)
Tube Damping - Actual Value (Page 176)
El.Dyn.Sens. - Actual Value (Page 177)
Freq.Fluct. - Actual Value (Page 178)
Tube Damp.Fluct. - Actual Value (Page 179)
Options:
OFF
ON
Factory setting:
ON
"Advanced Diagnostics" Transducer Block/base index 2200
Parameter
Write access
with operating
mode
(MODE_BLK)
Description