Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block
Endress+Hauser 173
Density -
Ref.Value Factory
read only The reference value for the density appears on the display.
Display:
5-digit floating-point number, incl. units
System Unit
Density
AUTO - OOS Use this parameter to select the unit for displaying the density.
The unit you select here is also valid for:
Simulation
Density response value for EPD
Density adjustment value
Options:
Metric g/cm
3
; g/cc; kg/dm
3
; kg/l; kg/m
3
; SD 4 °C, SD 15 °C,
SD 20 °C; SG 4 °C, SG 15 °C, SG 20 °C
US lb/ft
3
; lb/gal; lb/bbl (normal fluids); lb/bbl (beer);
lb/bbl (petrochemicals); lb/bbl (filling tanks)
Imperial lb/gal; lb/bbl (beer); lb/bbl (petrochemicals)
Factory setting:
kg/l
SD = Specific Density, SG = Specific Gravity
The specific density is the ratio of fluid density to water density (at water
temperature = 4, 15, 20 °C).
Density -
Ref.Value User
read only The reference value for the density appears on the display.
Display:
5-digit floating-point number, incl. units
Density -
Actual Value
read only The measured density appears on the display.
Display:
5-digit floating-point number, incl. units
Density - Min.Value read only The lowest density value since the saved values were last reset appears on
the display.
Display:
5-digit floating-point number, incl. units
Density - Max.Value read only The highest density value since the saved values were last reset appears on
the display.
Display:
5-digit floating-point number, incl. units
Density -
History 1…10
read only The last ten density values since the saved values were last reset appear on
the display.
Display:
5-digit floating-point number, incl. units
Density - Deviation read only This function displays the deviation between the measured density and the
reference values selected in the parameter "Adv. - Select Ref.Condition"
(FACTORY or USER)Page 170.
Display:
5-digit floating-point number, incl. units
"Advanced Diagnostics" Transducer Block/base index 2200
Parameter
Write access
with operating
mode
(MODE_BLK)
Description