55
CnF. 5CnF. 5
CnF. 5CnF. 5
CnF. 5
NOTENOTE
NOTENOTE
NOTE: when an input is programmed as RUN/
HOLD selector related with the status, the RUN/
HOLD selection by keyboard has no effect.
ru.St = Input contact used for RUN /HOLD selection.
The instrument toggles from one status to the
other one when the transition selected by
"d2.St" [C.H04] is detected.
NOTE:NOTE:
NOTE:NOTE:
NOTE: for more details about the RUN/HOLD
selection, see the related note at pag.54.
Pr.Ab = Input contact used for program abort (Abort
when the logic level is equal to "1").
NOTENOTE
NOTENOTE
NOTE: for ABORT function only, the input
status must be maintained for more than 3
seconds.
Pr.SL = Input contact used for program selection (for
more details see NOTE 2) of the "d1.Fn"
("Logic input 1 function" [C.H01]) parameter).
Au.ñA = Input contact used for Auto/Manual selection
(Manual when logic level is “1”)
O.LIñ = Input contact used for output limiter activation
(Output limited when logic level is “1”)
ñ.rSt = Input contact used to reset (acknowledge)
alarm (Reset when logic level is “1”)
rE.dr = Input contact used for Reverse/Direct control
action selection (Direct when logic level is “1”)
- Logic input DIG1 - contact status - [C.H02]- Logic input DIG1 - contact status - [C.H02]
- Logic input DIG1 - contact status - [C.H02]- Logic input DIG1 - contact status - [C.H02]
- Logic input DIG1 - contact status - [C.H02]
This parameter will be available only when "d1.Fn" [C.H01] is
different from “nonE”.
Range: CLSd = The input is at logic level “1” when contact is
closed.
When "d1.Fn" [C.H01] is equal to “ru.St”, the
instrument toggles from one status to the other
one when the transition from contact open to
contact close is detected.
OPEn = The input is at logic level “1” when contact is
open.
When "d1.Fn" [C.H01] is equal to “ru.St”, the
instrument toggles from one status to the other
one when the transition from contact close to
contact open is detected.
- Logic input DIG 2 function - [C.H03]- Logic input DIG 2 function - [C.H03]
- Logic input DIG 2 function - [C.H03]- Logic input DIG 2 function - [C.H03]
- Logic input DIG 2 function - [C.H03]
This parameter will be available only when input contact option
is fitted.
Range:nonE = Input contact not used
ru.SL = Input contact used for RUN /HOLD selection.
The status is related to the input level (RUN
when the logic level is "1").